Глюкометр Bionime Rightest GM 550

Test strip GS550 Coding Auto coding calibration Enzyme / Measurement technology GOD / Electrochemical sensor Blood sample source Capillary whole blood Minimum sample volume 0.75 microliters  (0.00075c.c.) HCT range 30 - 57% Measuring range 10 - 600mg/dL (0.6 - 33.3 mmol/L) Meter dimensions (L x W x H) 90.6 x 46.0 x 16.5 mm (3.56 x 1.81 x 0.64 inch) Display area (L x W) 46.6 x 33 mm (1.83 x 1.32 inch) Weight (with battery) 53.0 ± 5 g Power supply 2 CR2032 batteries Memory capacity 500 blood glucose test results with date and time Operating temperature 10 - 40°C (50 - 104°F) Operating relative humidity 10 - 90% Meter storage conditions -10 - 60°C (14 - 140°F) Test strip storage conditions 4 - 30°C (39 - 86°F), < 90% relative humidity Accuracy 95% of the measured glucose values shall fall within either
Glucose concentrations < 5.5mmol/L (100mg/dL), difference within ± 0.83mmol/L (15mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), bias within ± 15% Precision Glucose concentrations < 5.5mmol/L (100mg/dL), SD < 0.28mmol/L (5mg/dL)
Glucose concentrations ≥ 5.5mmol/L (100mg/dL), CV < 5%